Contact an Account ManagerDetermination of amoxicillin in human serum and plasma
Summary
Type: HPLC
Application #: 110510
Catalog#: 720149.40 - MN HPLC Column, Nucleosil C18, 120A, 3um, 4x100mm
Detection: Fluorescence, lex = 395 nm, lem = 485 nm
Conditions: Eluents: Methanesulfonic acid / acetonitrile 92:8 (v/v)
Temperature: 30 °C
Flow rate: 1.1 mL/min
Dervitisation: Fluorescamine (postcolumn)
Authors: Mascher H J
Source: J of Chromatography A 812 (1998) 221-226
Details
Legend: Chromatograms: (Amoxicillin: 2.7 min.)
A: Blank human plasma
B: Plasma sample of a volunteer 1 h (8.77 ug/ml) after oral intake of 500 mg
amoxicillin
C: Plasma sample of a volunteer 8 h (0.21 ug/m